Magnacoustics Inc™ © 2012 1-800-637-2282 REV-S 15 |
Genesis ULTRA Installation and Connections - continued
For Excite or newer systems go to step 13, For MR750 go to step 14
12: For Excite or newer systems locate Run 1085 labeled OW1-A21-J7 and remove the DB25
from J7 on the Wim Module and connect to the DB25 on the AutoVoice Breakout Box, then
connect “CB4” to the other DB25 on AutoVoice Breakout Box, and connect the other end of
“CB4” to J7 on the WIM Module.
13: For MR750 systems locate Run 3031 and remove the DB25 from J7 on the Audio Board in
the GOC and connect to the DB25 on the AutoVoice Breakout Box, then connect one side
of cable “CB4” to the other DB25 on AutoVoice Breakout Box, and connect the other end of
“CB4” to J7 on the Audio Board.
14: Connect the female side of cable “CB8” 6ft AC Adapter Cable to J12 on the rear of
Genesis Ultra module. Ensure that the power entry receptacle is set up for proper
voltage either 110 or 220 default is 110. CAUTION: Please ensure that all connections
are complete before connecting AC power cord.
15: Connect FM and optional AM Antenna if used to the Stereo. Please Reference Antenna
Connections in the Stereo Instructions for proper installation.
16: For GE MRI’s remove block off plate from J16, J17, J52 or J53, at penetration panel. All
others use supplied kit or fabricate a DB9 slot for filter as required.
17: Install RF filter with female toward computer room.
18: Locate existing electrical conduit between control console and computer room.
CAUTION: Assistance may be necessary to ensure Trauma free installation of cable CB1
19: Route and install Cable “CB1” (90 ft. DB9 Male to Female) with Female connector toward
Genesis Ultra module and Male toward penetration panel.
20: Connect “CB1” Male to the RF filter.
21: Zip tie cable “CB1” to existing supports, or cables from penetration panel to provide strain
relief.
22: Connect Female end of cable “CB1” to J11 on the rear of Genesis Ultra module.
23: From the scan room side of penetration panel, connect cable “CB2” (50ft. DB9 Female to
Female) to the RF filter you installed on penetration panel.
24: Zip tie cable “CB2” to existing cables from penetration panel to provide strain relief.
25: Route cable “CB2” on opposite side of the cold head (if used) to the front of the MRI
utilizing existing conduits.